The Pentacam® AXL is an enhanced version of the Pentacam® HR, which includes axial length measurement in addition to anterior segment tomography. Along with dependable Pentacam® software, this enables users to make accurate IOL calculations all with one device and one measurement process. The Pentacam® technology supports detection of diseases associated with the posterior corneal surface and makes for satisfied cataract patients. Detect abnormalities in patients’ eyes with the Fast Screening Report, comparing individual data to normative data Cataract-Pre-Op Display gives the necessary information to make the right IOL selection Minimizes transmission errors by transferring all relevant data to the IOL calculation software automatically Each result continues to optimize the IOL constants of your favored calculation formula for continuous improvement
Scheimpflug Camera Camera: digital CCD-Camera Light source: blue LED (475 nm UV-free) Processor: DSP with 400 Mio. operations/s Speed: 100 images in 2 seconds 1) Measurement range Axial length: 14 - 40 mm Curvature: 3 to 38 mm; 9 to 99 D Precision: ± 0,1 D Reproducibility: ± 0,1 D Operating distance: 80 mm (3.1 in) Technical specifications (Measurement head) Dimensions: (W x D x H) 270 x 208 x 430 mm (10.6 x 8.2 x 17.0 in) Weight: 7,9 kg (17.4 lbs) Voltage: 24 V DC Frequency: 50 - 60 Hz Max. power consumption: 37,4 W Recommended computer specifications: 3,4 GHz, 500 GB, Windows® 7, NVDIA NVS 315 graphic board