Terumo HX2 Heater Cooler Features The Terumo HX2 Heater Cooler Temperature Management System offers precise temperature control and many unique features including drip-proof self-sealing water fittings. Dual Channel - Two independent channels suppy temperature-controlled water to arterial and cardioplegia heat exchanges. Temperature Control - Each touch pad precisely adjusts the mix of water flow between cooling and heating sources in 1° C increments. Prime Mode - Automatically removes air from the internal and external circulation resulting in fast set-up times. Tank Divider - Partitioning the ice supply maximizes the cooling capability of a channel Tank Drain - Drains water from the tank during use without interrupting the water supply to the heat exchangers. Reduced Footprint - Small size improves accessibility to controls and eliminates clutter. Terumo HX2 Heater Cooler Specifications Physical Specifications Height: 35.5” (90 cm) Width: 20.4” (52 cm) Depth: 27.8” (71 cm) Weight (Tank empty): 188 lbs (85 kg) Weight (Tank full): 276 lbs (125 kg) Capacity: Water 40 liters (approx.) System Preformance Outlet water temperature setpoint control range: 1-42°C Outlet water temperature setpoint resolution: 1°C Outlet water temperature setpoint control accuracy: +/- 1°C Outlet water temperature display range: 0 - 50°C Outlet water temperature display resolution: 0.1°C Output water flow maximum at zero head: 25 L/min Maximum outlet pressure: 13 PSI