The Topcon 3D OCT-1 Maestro from EMS provides ophthalmologists with an advanced automated optical coherence tomography system and fundus camera for diagnosing optical pathologies. The full color fundus camera offers both 2D and 3D images in high resolution, while the Spectral Domain OCT provides 50,000 A-scans per second. The Topcon 3D OCT-1 Maestro also features a number of fundus image utilities, including Mosaic, stereo, patient education, measurement, and more. With automatic disc segmentation and auto-centering capabilities, as well as automatic layer segmentation analysis, this OCT enables precise diagnostics and detailed interactive reports. The instrument is compact and fully networkable for seamless integration within any size ophthalmology practice. Automatic alignment, focus, and capture Automatic disc segmentation 2D and 3D fundus images Wide-field OCT scan Cataract mode Color fundus camera with Live Fundus View 50,000 A-Scans per second Automatic centering of fovea and disc Reference database, interactive reports, and pre-defined reports Fundus image utilities: Mosaic, stereo, patient education, measurement, and more Automatic layer segmentation analysis of: Total retina, RNFL, GCL + IPL, GCL + IPL +RNFL Fully networkable with DICOM connectivity Rotatable touch-screen interface Customizable with the following options: Standard OCT, Anterior OCT, True Color Fundus, and Red Free Fundus
Observation & Photography of Fundus Image Scan Mode: Color, Red-free (display digital red-free) Picture Angle: 45°/30° or equivalent (digital zoom) Operating Distance 8 mm (in fundus photography) 6 mm (in anterior segment photography) Photographable Diameter of Pupil 45°: Ø4.0 mm or more Small pupil diameter: Ø3.3 mm or more Observation & Photographing of the Fundus/Anterior Segment Tomogram Scan Range On fundus Horizontal direction 3 – 12 mm Vertical direction 3 – 9 mm On cornea Horizontal direction 3 – 6 mm Vertical direction 3 – 6 mm Scan Speed: 50,000 A-Scans per second Lateral Resolution: 20 µm In-depth Resolution:6 µm Photographable Diameter of Pupil: Ø2.5 mm or more Internal Fixation Target: Dot matrix type organic EL (The display position can be changed and adjusted. The presenting method can be changed.) Electric Rating Source Voltage: AC 100 – 240V Power Input: 70 – 150VA Frequency: 50HZ – 60HZ Dimensions & Weight Dimensions: 1” – 17.4” (W) x 18.5” – 26.3” (D) x 20.4” – 28.4” (H) Weight: 3 lbs.